Chapter 9 SCXI—Signal Conditioning

Binary Reading

Ideal Reading

Actual Reading

Gain Error

Vos

Actual Voltage

in Binary Representation

Binary Offset

Figure 9-15. Ideal versus Actual Reading

Figure 9-15 shows the difference between the ideal reading and the actual reading. This difference is called Vos, or the binary offset, before the two readings intersect. The difference in slope between the actual and ideal readings is called the gain error.

One-point calibration removes the Vos (binary offset) by measuring a 0 V signal and comparing the actual reading to it. Two-point calibration removes the Vos (binary offset) and corrects gain error by first performing a one-point calibration. Then you measure a voltage at x volts and compare it to the actual reading. The x must be as close as possible to the full-scale range. The following sections explain how to perform a one-point and two-point calibration.

LabVIEW Measurements Manual

9-38