Chapter 6

Analog Input

instrumentation amplifier. For this device, the pin labeled AIGND (the analog input ground) is the measurement system ground.

MUX

CH0+

CH2+

CH1+

 

Instrumentation Amplifier

CH7+

+

MUX

+

CH0–

Vm

CH2–

CH1–

CH7–

AIGND

Figure 6-7. 8-Channel Differential Measurement System

© National Instruments Corporation

6-9

LabVIEW Measurements Manual

Chapter 6

Analog Input

In general, a differential measurement system is preferable because it rejects not only ground loop-induced errors, but also the noise picked up in the environment to a certain degree. Use differential measurement systems when all input signals meet the following criteria:

Low-level signals (for example, less than 1 V)

Long or non-shielded cabling/wiring traveling through a noisy environment

Any of the input signals require a separate ground-reference point or return signal

An ideal differential measurement system reads only the potential difference between its two terminals—the positive (+) and negative (–) inputs. Any voltage present at the instrumentation amplifier inputs with respect to the amplifier ground is called a common-mode voltage. An ideal differential measurement system completely rejects (does not measure) common-mode voltage, as shown in Figure 6-8.

 

 

 

 

Instrumentation Amplifier

 

 

 

+

 

 

 

 

 

 

 

Grounded Signal Source

+

 

 

 

 

 

Vs

 

+

 

 

 

 

 

Measured Voltage

 

 

 

 

 

Common Mode Voltage,

+

 

 

 

Vcm

 

 

 

Ground Potential, Noise, etc.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 6-8. Common-Mode Voltage

LabVIEW Measurements Manual

6-10

www.ni.com

Chapter 6

Analog Input

Referenced Single-Ended Measurement System

An RSE measurement system is used to measure a floating signal, because it grounds the signal with respect to building ground. Figure 6-9 depicts a 16-channel RSE measurement system. You should use this measurement system only when you need a single-ended system and your device does not work with NRSE measurement.

MUX

CH0

CH2

CH1

Instrumentation Amplifier

+

+

CH15

Vm

AIGND

Figure 6-9. 16-Channel RSE Measurement System

© National Instruments Corporation

6-11

LabVIEW Measurements Manual

Chapter 6

Analog Input

Nonreferenced Single-Ended Measurement System

DAQ devices often use a variant of the RSE measurement technique, known as the NRSE measurement system. In an NRSE measurement system, all measurements are made with respect to a common reference, because all of the input signals are already grounded. Figure 6-10 depicts an NRSE measurement system where AISENSE is the common reference for taking measurements and AIGND is the system ground. All signals must share a common reference at AISENSE.

MUX

CH0+

CH1+

CH2+

Instrumentation Amplifier

+

+

CH15+

Vm

AISENSE

AIGND

Figure 6-10. 16-Channel NRSE Measurement System

In general, a differential measurement system is preferable because it rejects not only ground loop-induced errors, but also the noise picked up in the environment to a certain degree. On the other hand, the single-ended configuration allows for twice as many measurement channels and is acceptable when the magnitude of the induced errors is smaller than the required accuracy of the data.

LabVIEW Measurements Manual

6-12

www.ni.com