Chapter 9 SCXI—Signal Conditioning

Figure 9-1 shows some common types of transducers/signals and the required signal conditioning for each.

 

Transducers/Signals

 

 

 

Signal Conditioning

 

 

 

 

 

 

 

 

 

 

 

 

 

Thermocouples

 

 

 

Amplification, Linearization, and

 

 

 

 

 

 

 

Cold-Junction Compensation

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Current Excitation, Four Wire

 

 

 

 

RTDs

 

 

 

and Three Wire Configuration,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Linearization

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Strain Gauges

 

 

 

Voltage Excitation Bridge

 

 

 

 

 

 

 

Configuration, and Linearization

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Common Mode

 

 

 

Isolation Amplifiers

 

 

DAQ Device

 

or High Voltages

 

 

 

(Optical Isolation)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Loads Requiring AC Switching

 

 

 

Electromechanical Relays

 

 

 

 

or Large Current Flow

 

 

 

or Solid-State Relays

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Signals with High

 

 

 

Lowpass Filters

 

 

 

 

Frequency Noise

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 9-1. Common Types of Transducers/Signals and Signal Conditioning

Amplification

The most common type of signal conditioning is amplification. Amplify electrical signals to improve the digitized signal accuracy and to reduce noise.

For the highest possible accuracy, amplify the signal so the maximum voltage swing equals the maximum input range of the ADC, or digitizer. Your system should amplify low-level signals at the DAQ device or at the SCXI module located nearest to the signal source, as shown in Figure 9-2.

© National Instruments Corporation

9-3

LabVIEW Measurements Manual